The following questions may be often asked about autotransfusion, and the answers are provided here for your convenience. If you have further questions, please contact our 24-hour technical support.

  1. During rapid blood loss conditions, is it normal for the filter chamber in the Ocean 2050, Oasis 3650, or Express 4050 ATS chamber models to fill with clot?

  2. In the event that the filter chamber does fill up with clot and drainage starts to spill over into the ATS chamber, can I continue reinfusion?

  3. What should I do if the patient’s rate of bleeding exceeds the maximum infusion pump rate setting?

  4. What should I do when an “air-in-line” pump alarm sounds or when an “occlusion” alarms sounds?

  5. What should I do if the 2450 self-filling ATS bag does not fill completely after activating blood bag?

  6. In the event blood reinfusion does not appear to be flowing freely into the patient during gravity (non-pressure) ATS bag reinfusion, what should I check for?

  7. What do I do when clotting within the ATS bag occurs during patient reinfusion?

  8. How long can I collect blood before I reinfuse the patient?

  9. Do the Atrium Self-Filling 2450 and In-Line 2550 Blood Bags contain any latex?

1. During rapid blood loss conditions, is it normal for the filter chamber in the Ocean 2050, Oasis 3650, or Express 4050 ATS chamber models to fill with clot?
When rapid or massive blood loss is encountered, blood may not be allowed to pool in the chest long enough to become defribrinated. In general, the more rapid the bleeding, the more clotting you can expect to see. The purpose of the filter chamber is to screen large blood particulates during collection. The filter chamber has a large 300ml filtering capacity and is calibrated to give you useful diagnostic information about your patient’s condition, i.e., is there a lot of clotting or a little? An important safety feature of Atrium’s advanced filter chamber design is that it can allow drainage to spill over in the rare event that the filter chamber would fill up or clot off, preventing any possible fluid to backup the patient tube. The Atrium system provides unrestricted patient drainage under all patient condition, thus helping to reduce the risk of cardiac tamponade. When rapid or massive blood loss can be expected, an effective dose of “citrate” anticoagulant is recommended to help minimize clotting in the chest drain during such conditions. Return to Top

2. In the event that the filter chamber does fill up with clot and drainage starts to spill over into the ATS chamber, can I continue reinfusion?
Yes. If the blood is considered suitable for reinfusion you may continue ATS. The spill over safety feature of the filter chamber does not preclude ATS because a microemboli blood filter, which must be used during reinfusion, will screen out any harmful particulates to the patient. However, during episodes of heavy blood loss, you can anticipate a fair amount of clotting and accessing the blood for subsequent reinfusion from the ATS chamber or ATS bag may be difficult. If rapid or massive blood loss can be expected, an effective dose of “citrate” anticoagulant is recommended to help minimize clotting and enhance ATS efficiency during such conditions. Return to Top

3. What should I do if the patient’s rate of bleeding exceeds the maximum infusion pump rate setting?
For clinical situations where blood loss is massive or for emergency chest traums, use of either an Atrium 2450 self-filling ATS bag or 2550 in-line ATS bag will provide a more rapid infusion technique as compared with a continuous infusion pump method. Use of multiple ATS blood bags will accommodate larger blood volumes, faster than an infusion pump set at a maximum infusion rate. The rate of gravity reinfusion from an ATS bag is equal to or faster than many infusion pumps. Hence, use of a flexible ATS system such as Atrium may provide the most efficient means for conducting hospital-wide chest drainage ATS. Return to Top

4. What should I do:
• When an “air-in-line” pump alarm sounds?
Make sure that you are using a non-vented blood administration set. Check to see that all connections are air-tight including the filter spike and any luer-lock connections. Failure to position drip chamber in a spike up position may result in an “air-in-line” pump alarm. An alarm may also result if the “total volume to be reinfused” setting on the pump exceeds the total volume of blood in the Atrium chest drain. After you locate and correct the source of the air in the IV circuit, commence repriming until all in-line air is fully purged from the IV circuit and reset the infusion pump for continuous autotransfusion. Failure to purge all air from the complete IV circuit can result in an air emboli.

• When an “occlusion” alarms sounds?
First, check to see that all clamps are open and there are no kinks in the tubing. During continuous ATS, the infusion pump is “pulling” against both negative pressure in the chest drain and gravity. To minimize the force of gravity, the infusion pump should be positioned low on the IV pole, at approximately the same level as the drain. It is also recommended to set the pressure (psi) setting on the pump to “maximum” or “high” setting. If these steps are taken and the occlusion alarm persists, it may indicate that clotting has occurred somewhere in the IV circuit, i.e., the ATS sump port, ATS access line, microemboli blood filter, or IV blood set. Under certain conditions, shed blood has been reported to clot during ATS collection which requires a medical decision whether to continue or discontinue autotransfusion, or add an anticoagulant such as ACD-A or CPD, to reinstitute patient reinfusion. Return to Top

5. What should I do if the 2450 self-filling ATS bag does not fill completely after activating blood bag?
Any remaining air in the ATS bag can be easily displaced back into the chest drain by gently squeezing the ATS bag. This procedure is quite normal and will allow more blood volume to transfer from the ATS chamber directly into the self-filling ATS bag. Repeat as often as necessary until all air is displaced and the ATS chamber volume has been transferred into the ATS bag. Return to Top

6. In the event blood reinfusion does not appear to be flowing freely into the patient during gravity (non-pressure) ATS bag reinfusion, what should I check for?
Make sure the filtered air vent located on top of the Atrium ATS bag and all IV line clamps are open to maximize flow to the patient. It is also important to periodically check the ATS bag during patient reinfusion for any signs of intra-ATS bag clotting by gently tipping the ATS bag to one side to observe any clot formation.

To troubleshoot this situation, first check to be sure that all connections are air-tight. Inspect the suction tubing and connections for possible cracks, leaks, kinks, or occlusions. You may need to simply bypass a “leaky” wall canister. Try connecting the chest drain to a different suction source or wall regulator. When multiple chest drains are “Y” connected to a single suction source, if possible, reconnect the drains to separate suction sources. Finally, replace the chest drain if you suspect the unit is cracked or damaged.
Return to Top

7. What do I do when clotting within the ATS bag occurs during patient reinfusion?
Under certain conditions, shed blood has been reported to clot during ATS collection. If you observe clot formation in the ATS bag during reinfusion, unfortunately, it may be too late to do anything about it. Adding “citrate” ACD-A anticoagulant to the bag will reduce any further clotting, however, it will not lyse, or break-up, clots which have already formed. In this situation, you can try gently manipulating, or tipping, the ATS bag to one side. This will often move clot formation away from the spike port area and improve blood reinfusion to the patient. Return to Top

8. How long can I collect blood before I reinfuse the patient?
The American Association of Blood Banks (AABB) dictates that blood no older than 6 hours can be reinfused back to the patient. Within this time constraint, is the ability to collect blood for 1 hour then reinfuse over 5 hours (for a total of 6), collect blood for 2 hours then reinfuse over 4 hours (for a total of 6), etc. Return to Top

9. Do the Atrium Self-Filling 2450 and In-Line 2550 Blood Bags contain any latex?
The 2450 and 2550 do not contain natural rubber latex (NRL) or dry natural rubber (DNR) and bears labeling with the ‘LF’ symbol and the statement “Latex Free”. Return to Top