The following questions may be often asked about the Express drain, and the answers are provided here for your convenience. If you have further questions, please contact our 24-hour technical support.

  1. Is it required to fill the air leak monitor?

  2. What does it mean when the vacuum indicator is visible?

  3. When will I see a rise in the air leak monitor column?

  4. Should the manual vent be used during gravity drainage?

  5. What should I do when the suction monitor bellows is not expanded to the [delta symbol] mark when the regulator is set at -20cmH2O or higher?

  6. What should I do when the bellows does not fully expand to the [delta symbol] mark after I increase the suction source vacuum?

  7. Does the bellows need to expand beyond the [delta symbol] mark for a -10cmH2O regulator setting?

  8. How do I confirm my patient has a pneumothorax?

  9. What does it mean when the small float ball is located at the bottom of the air leak monitor?

  10. What should I do if the chest drain gets knocked over?

  11. If the regulator of the Express is set at -20cmH2O (preset level), how do I know my patient is experiencing -20cmH2O?

  12. Does the Express drain contain any latex?

  13. Is the Express drain MRI compatible?

  14. How do I dispose of the system?

1. Is it required to fill the air leak monitor?
It is not required to fill the air leak monitor for seal protection. However, the air leak monitor should be filled for confirmation and detection of air leaks.
Return to Top

2. What does it mean when the vacuum indicator is visible?
The [check mark symbol] indicates that there is vacuum present in the system (negative pressure, i.e. -20cmH2O). When vacuum is not present (atmospheric pressure) no symbol will appear. All patient tube connections and the vacuum indicator window should be checked regularly for vacuum confirmation. Return to Top

3. When will I see a rise in the air leak monitor column?
A rise in the air leak monitor column will only be seen if there is an increase in negative pressure on the patient side. When changing suction pressure from a higher level to lower level, depress the manual high negativity vent to reduce excess vacuum to the lower prescribed level. Return to Top

4. Should the manual vent be used during gravity drainage?
No. It is not recommended to depress the manual vent during gravity (no suction) drainage. Depressing the manual vent during gravity drainage could potentially result in positive pressure (atmospheric air) getting back to the patient. Return to Top

5. What should I do when the suction monitor bellows is not expanded to the [delta symbol] mark when the regulator is set at -20cmH2O or higher?
The position of the bellows across the suction monitor window will alert the operator that the suction source has fallen below the minimum operating range for the prescribed suction control setting. Simply increase the vacuum source to -80mmHg or higher. The suction monitor bellows must expand to the [delta symbol] mark or beyond for a -20cmH2O or higher suction regulator setting. Return to Top

6. What should I do when the bellows does not fully expand to the [delta symbol] mark after I increase the suction source vacuum?
Dry suction chest drains require higher levels of vacuum pressure and air flow from the suction source to operate efficiently at each suction control setting as compared to traditional water controlled operating systems. The suction source should provide a minimum vacuum pressure of -80mmHg at 20 liters of air flow per minute for chest drain operating efficiency at a suction control setting of -20cmH2O. The suction source should be greater than -80mmHg when multiple chest drains are connected to a single suction source. If the bellows does not fully expand to the [delta symbol] mark, it may simply be that the suction source is not functioning to its full potential to provide the minimum vacuum pressure or air flow required to “drive” the suction control regulator. Additionally, conditions may exist that can reduce, or “restrict” air flow from the suction source. A restrictive clamp, connector, or kink in the suction line tubing can potentially “starve” the chest drain of air flow. A leak in a connection or wall canister, along with extensive lengths of suction tubing can also reduce air flow to the unit.

To troubleshoot this situation, first check to be sure that all connections are air-tight. Inspect the suction tubing and connections for possible cracks, leaks, kinks, or occlusions. You may need to simply bypass a “leaky” wall canister. Try connecting the chest drain to a different suction source or wall regulator. When multiple chest drains are “Y” connected to a single suction source, if possible, reconnect the drains to separate suction sources. Finally, replace the chest drain if you suspect the unit is cracked or damaged.
Return to Top

7. Does the bellows need to expand beyond the [delta symbol] mark for a -10cmH2O regulator setting?
No. For a regulator setting less than -20cmH2O suction (-10cmH2O), any observed bellows expansion across the monitor window will confirm suction operation. The bellows need not be expanded to the [delta symbol] mark for suction pressures less than -20cmH2O, just visibly expanded to confirm suction operation. Return to Top

8. How do I confirm my patient has a pneumothorax?
If there are no air bubbles observed going from right to left in the water seal, there is no patient air leak. If bubbling is present (constant or intermittent) in the water seal air leak monitor, this will confirm an air leak is present. In order to confirm that your patient’s chest catheter(s) are patent, temporarily turn suction off and check for oscillation of the patient pressure float ball in the water seal column coinciding with patient respiration.
Return to Top

9. What does it mean when the small float ball is located at the bottom of the air leak monitor?
If the small float ball is located and oscillating at the bottom of the air leak monitor with no bubbling, there is no apparent patient air leak. However, the air leak monitor should be carefully monitored for the presence of an occasional or intermittent air leak.
Return to Top

10. What should I do if the chest drain gets knocked over?
We recommend that the drain be placed back into the upright position; however it will continue operating if knocked over. After the drain is set upright, check the fluid level in the air leak monitor. Atrium provided a convenient needleless luer port on the back of the Express drain to adjust the fluid level. Simply use a syringe (no needle) to add or remove fluid in the air leak monitor, if required.
Return to Top

11. If the regulator of the Express is set at -20cmH2O (preset level), how do I know my patient is experiencing -20cmH2O?
With the dry suction regulator there is a small amount of variance in suction at any given suction level. That variance is +/- 2cmH2O. That means if the regulator is set at -20cmH2O then the patient could be experiencing anywhere from -18cmH2O to -22cmH2O. That is a small amount of variance and not clinically dangerous for the patient. Atrium’s dry suction regulator are pre-assembled and tested before being added to each drain.
Return to Top

12. Does the Express drain contain any latex?
The Express series chest drains do not contain natural rubber latex (NRL) or dry natural rubber (DNR) and bear labeling with the ‘LF’ symbol and the statement “Latex Free”. Return to Top

13. Is the Express drain MRI compatible?
The metal contained within the drain (small spring for manual vent) is stainless steel which is non-magnetic. The metallic material used in Atrium Medical Corporation’s chest drainage units will have no effect on the MRI equipment. This will allow for the product to remain attached to the patient while the MRI is being performed. Return to Top

14. How do I dispose of the system?
Disposal of system and contents must be in accordance with approved hospital infection control standards. To minimize the possibility of fluid leaking from the drain, it is suggested that the patient line occluded using the blue slide clamp provided with each drain. Return to Top